UJI KESESUAIAN BERKAS CAHAYA KOLIMATOR PADA PESAWAT KONVENSIONAL DI INSTALASI RADIOLOGI RUMAH SAKIT ISLAM IBNU SINA PEKANBARU

Nugroho, Hadits (2024) UJI KESESUAIAN BERKAS CAHAYA KOLIMATOR PADA PESAWAT KONVENSIONAL DI INSTALASI RADIOLOGI RUMAH SAKIT ISLAM IBNU SINA PEKANBARU. D3 Teknik Radiologi thesis, Universitas Awal Bros.

[img] Text
KTI HADITS CHAIRUNNISA_21002021.pdf

Download (2MB)
[img] Text
DAFTAR PUSTAKA HADITS CHAIRUNNISA.pdf

Download (93kB)
[img] Text
BAB I Hadits Chairunnisa.docx

Download (30kB)

Abstract

ABSTRAK Uji Kesesuaian adalah pengujian untuk memastikan bahwa pesawat sinar- X memenuhi persyaratan keselamatan radiasi dan memberikan informasi diagnosa atau pelaksanaan radiologi yang tepat dan akurat. Salah satu uji yang dilakukan yaitu uji kesesuaian luas berkas Cahaya kolimator dengan berkas sinar-X dan uji titik focus. Di Instalasi Radiologi Rumah Sakit Islam Ibnu Sina Pekanbaru terakhir pengujian pada Agustus 2023 sedangkan berdasarkan KEMENKES RI No.1250 tahun 2009 tentang uji kesesuaian berkas cahaya kolimator bahwasannya dilakukan pengujian dalam satu bulan sekali. Tujuan dilakukan pengujian ini untuk mengetahui hasil uji kesesuaian berkas cahaya kolimator dan titik fokus masih dalam batas toleransi yang telah ditetapkan. Jenis penelitian yang digunakan metode Kuantitatif Deskriptif dengan pendekatan Observasional, dengan cara pengambilan data menggunakan collimator test tool dan beam aligment dan dilakukan pengujian pada bulan Mei 2024 di Instalasi Radiologi Rumah Sakit Islam Ibnu Sina Pekanbaru. Berdasarkan hasil uji diperoleh nilai rata-rata sumbu X yaitu -0,561 cm dan sumbu Y -0,2735 cm untuk focal spot kecil, untuk focal spot besar diperoleh nilai rata-rata sumbu X -0,892 cm dan sumbu Y -0,14 cm dari hasil tersebut ≤ 2% dari FFD yg telah ditetapkan. Untuk uji ketetapan titik fokus pada empat kali percobaan tidak ada penyimpangan titik pusat yang melebihi batas toleransi. Kata kunci: Focus Film Distance, Focal Spot ABSTRACT Conformity Test is a test to ensure that the X-ray aircraft meets radiation safety requirements and provides precise and accurate diagnostic or radiological information. One of the tests carried out is the suitability test of the area of the collimator light beam with the X-ray beam and the focus point test. At the Radiology Installation at Ibnu Sina Islamic Hospital Pekanbaru, the last inspection was carried out in August 2023, while based on the Regulation of the Minister of Health of the Republic of Indonesia Number 1250 of 2009 concerning Collimator Beam Conformity Tests, testing is carried out once a month. The purpose of this test is to determine whether the test results for the suitability of the collimator light beam and the focus point are still within the specified tolerance limits. The type of research used is the Quantitative Descriptive method with an Observational approach, with data collection using collimator and beam alignment test equipment and testing carried out in May 2024 at the Radiology Installation at the Ibnu Sina Islamic Hospital Pekanbaru. Based on the test results, the average value of the X axis is -0.561 cm and the Y axis is -0.2735 cm for small focal spots, for large focal spots the average value of This result is ≤ 2% of the predetermined FFD. For the focus point fixation test, in four trials there was no center point deviation that exceeded the tolerance limit. Keywords: Focus Film Distance, Focal Spot

Item Type: Thesis (D3 Teknik Radiologi)
Subjects: D3 Radiologi > X-Ray Konvensional
Divisions: Fakultas Ilmu Kesehatan > D3 Radiologi
Depositing User: Hadits Chairunnisa Fadillah Nugroho
Date Deposited: 27 Feb 2025 02:38
Last Modified: 27 Feb 2025 02:38
URI: http://repository.univawalbros.ac.id/id/eprint/378